Quantum Cascade Laser Analyser for Greenhouse Gases and Air Pollutants
Laboratory
optical and precision equipments (excl. glasses)
UK
Published value
Not published
Submission deadline Not published
Lots published1
Procurement Executive Summary
AI & Search Synopsis
Generated from official OCDS record
Tenderline Synopsis: National Physical Laboratory: "Quantum Cascade Laser Analyser for Greenhouse Gases and Air Pollutants". Published status: complete. Published value: Value not published. 1 published lot. Submission deadline not published. See the official notice for participation instructions.
| Contracting Authority | National Physical Laboratory | Scope & Categories | Not published | Submission Window | complete No deadline published |
|---|---|---|---|---|---|
| Submission Gateway | Direct notice route | Legal Basis & Regime | Standard procurement | Estimated Value (exc. VAT) | Not published |
Bidder Intelligence · Authority Profile: National Physical Laboratory
Market Analytics
Derived from OCDS awards & bid statistics
Published history for National Physical Laboratory. These figures describe retained records, not a forecast of bids or a measure of buyer bias.
Published-to-award variance
Not availableInsufficient comparable data
Competition Density
3.5Bids / Report
Supplier ConcentrationNo estimate
Insufficient attributable awardsNo concentration estimate availablePayment Terms
Check noticePublished terms
Coverage: 59 active published awards; 23 bid reports (which may be per lot). Supplier values exclude multi-supplier awards, frameworks and DPS, and use GBP only. They are published award values, not payments. Published-to-award variance compares single-lot, single-award, single-supplier GBP procedures with explicitly non-framework/non-DPS status; increases remain in the average. Unpublished data stays unknown. Awarded suppliers are winners, not all bidders.
Procedure terms
Contracting AuthorityNational Physical Laboratory | Procedure methodNot published | Procurement categoryNot published |
Statuscomplete | Framework / DPSNot published | CompetitionNot published |
Above thresholdNot published | Legal basisNot published | Tender period startsNot published |
Clarification deadlineNot published | Electronic submissionNot published | Submission languagesNot published |
Published28 Jul 2023, 10:25 BST | Last source update28 Jul 2023, 10:25 BST | Recurring procurementNot published |
ClassificationLaboratory, optical and precision equipments (excl. glasses) | ||
Delivery area UK | ||
OCIDocds-h6vhtk-03eb22 | ||
What is being bought
A rack mounted quantum cascade laser instrument capable of high precision (< 0.5 nmol mol-1 for all except carbon dioxide ≤ 25 nmol mol-1) simultaneous measurements of carbon dioxide, methane nitrous oxide, carbon monoxide, ammonia, nitrogen monoxide, nitrogen dioxide and sulfur dioxide in the 0 – 10 µmol mol-1 range (0 – 30 µmol mol-1 for methane and 0- 5000 µmol mol-1 for carbon dioxide).
What changed
From the official release history
- Status changed to complete
28 Jul 2023, 10:25 BST - Official notice release published
28 Jul 2023, 10:25 BST - Buyer information updated
28 Jul 2023, 10:25 BST
Lots and requirements (1)
Published by the contracting authority
- Lot 1 · #1Individual lot title not publishedStatus not publishedPublished valueNot publishedMulticomponent quantum cascade laser analyser with ≥ 1 Hz data acquisition rate that allows the simultaneous measurement of carbon dioxide (CO2), methane (CH4), nitrous oxide (N2O), carbon monoxide (CO), ammonia (NH3), nitrogen monoxide (NO), nitrogen dioxide (NO2), sulphur dioxide (SO2) with the following amount fractions ranges and one sigma (1σ) precisions: Carbon dioxide (amount fraction range: 0 – 5000 µmol mol-1; precision 25 nmol mol-1), methane (amount fraction range: 0 – 30 µmol mol-1; precision 0.22 nmol mol-1), nitrous oxide (amount fraction range: 0 – 10 µmol mol-1; precision 0.1 nmol mol-1), carbon monoxide (amount fraction range: 0 – 10 µmol mol-1; precision 0.01 nmol mol-1), ammonia (amount fraction range: 0 – 10 µmol mol-1; precision 0.5 nmol mol-1), nitrogen monoxide (amount fraction range: 0 – 10 µmol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater), nitrogen dioxide (amount fraction range: 0 – 10 µmol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater) and sulphur dioxide (amount fraction range: 0 – 10 µmol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater). The instrument shall have a flow rate of 0.5 – 1.5 slpm, a leak rate of ≤ 10¯⁹ mbar xl/s and response time of < 2 minutes. It shall be able to operate under environmental conditions of ambient temperature: 15⁰C to 30⁰C and ambient humidity: 30-80% RH without condensing. The wetted materials shall be corrosion resistance preferable SS 316/316L. A gas connection of ¼” or 6mm Swagelok is required with an inlet pressure of 1 – 5 bar absolute. The instrument shall have the following connectivity: 100-1000 Base-T Ethernet, USB 2.0 or better, serial RS232/485, Modbus TCP and an isolated 4-20 mA or 0-10 Vdc analogue output. Power consumption ideally should < 500 W and with a total weight of < 50 kg. The analyser shall be CE/UKCA certificated with a minimum of 12-month warranty.Contract periodNot publishedEligibilityNot publishedOptions / renewalNot published
Timeline
- Procedure published
28 Jul 2023, 10:25 BST - Award active
Not published · Not published - Contract active
Signed 28 Jul 2023, 00:00 BST · CHF 205,274
Commercial outcome and competition
Awards Miro Analytical AG Not published · Not published · active |
Contracts Quantum Cascade Laser Analyser for Greenhouse Gases and Air Pollutants CHF 205,274 · signed 28 Jul 2023, 00:00 BST · active |
Bid statisticsNo aggregate bid statistics published |
Buyer and organisations in this procedure
National Physical Laboratory
Contracting authority GB-FTS-155Documents (0)
Official links; attachments are not copied
No linked documents are published
Related procedures (0)
No related procedures published
Planning and rationale
Planning budgetNot published |
No-engagement rationaleNot published |
Procedure rationaleMiro Analytical AG is the only vendor that can provide an analytical solution that enables the simultaneous measurement of all the major greenhouse gases (CO2, CH4, N2O and CO) and air quality pollutants (NO, NO2, SO2 and NH3) with the required high precision and broad amount fraction ranges that are required.
Unique features include:
1. The ability to measure all these compounds simultaneously due to a unique dual colour laser system.
2. The achieved precisions with this analyser match or exceed other optical spectrometers that are typically measuring one or two compounds.
3. The range of amount fractions that are possible to measure exceed those typical of other optical spectrometers that are measuring one or two compounds. |