What is being bought
Published requirement descriptionA rack mounted quantum cascade laser instrument capable of high precision (< 0.5 nmol mol-1 for all except carbon dioxide ≤ 25 nmol mol-1) simultaneous measurements of carbon dioxide, methane nitrous oxide, carbon monoxide, ammonia, nitrogen monoxide, nitrogen dioxide and sulfur dioxide in the 0 – 10 µmol mol-1 range (0 – 30 µmol mol-1 for methane and 0- 5000 µmol mol-1 for carbon dioxide).
CPV: Laboratory, optical and precision equipments (excl. glasses)
The lotProcurement structure
single lot · no lot value publishedLot #1Multicomponent quantum cascade laser analyser with ≥ 1 Hz data acquisition rate that…
Multicomponent quantum cascade laser analyser with ≥ 1 Hz data acquisition rate that allows the simultaneous measurement of carbon dioxide (CO2), methane (CH4), nitrous oxide (N2O), carbon monoxide (CO), ammonia (NH3), nitrogen monoxide (NO), nitrogen dioxide (NO2), sulphur dioxide (SO2) with the following amount fractions ranges and one sigma (1σ) precisions:
Carbon dioxide (amount fraction range: 0 – 5000 µmol mol-1; precision 25 nmol mol-1), methane (amount fraction range: 0 – 30 µmol mol-1; precision 0.22 nmol mol-1), nitrous oxide (amount fraction range: 0 – 10 µmol mol-1; precision 0.1 nmol mol-1), carbon monoxide (amount fraction range: 0 – 10 µmol mol-1; precision 0.01 nmol mol-1), ammonia (amount fraction range: 0 – 10 µmol mol-1; precision 0.5 nmol mol-1), nitrogen monoxide (amount fraction range: 0 – 10 µmol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater), nitrogen dioxide (amount fraction range: 0 – 10 µmol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater) and sulphur dioxide (amount fraction range: 0 – 10 µmol mol-1; precision 1 % or 1 nmol mol-1, whichever is greater).
The instrument shall have a flow rate of 0.5 – 1.5 slpm, a leak rate of ≤ 10¯⁹ mbar xl/s and response time of < 2 minutes. It shall be able to operate under environmental conditions of ambient temperature: 15⁰C to 30⁰C and ambient humidity: 30-80% RH without condensing. The wetted materials shall be corrosion resistance preferable SS 316/316L. A gas connection of ¼” or 6mm Swagelok is required with an inlet pressure of 1 – 5 bar absolute. The instrument shall have the following connectivity: 100-1000 Base-T Ethernet, USB 2.0 or better, serial RS232/485, Modbus TCP and an isolated 4-20 mA or 0-10 Vdc analogue output. Power consumption ideally should < 500 W and with a total weight of < 50 kg. The analyser shall be CE/UKCA certificated with a minimum of 12-month warranty.
Procedure terms & legal framework
From the official record- Procedure method
- Award procedure without prior publication of a call for competition
- Legal basis
- CELEX · 32014L0024 Public Contracts Regulations 2015 (Directive 2014/24/EU)
- Procurement category
- Goods
- Notice status
- Awarded without competition
- Procurement threshold
- Unspecified in notice
- Regulatory regime
- Standard
- Commercial structure
- Stand-alone contract (no framework)
- Competition type
- Direct award (no call for competition)
- GPA / WTO covered
- Unspecified in notice
- Recurring procurement
- Unspecified in notice
- First published
- 28 Jul 2023, 10:25 BST
- Last source update
- 28 Jul 2023, 10:25 BST
- FTS notice number
2023/S 000-021940- Classification (CPV)
- Delivery area
- Official registry OCID
ocds-h6vhtk-03eb22Find a Tender
- Why no open competition
- The procurement falls outside the scope of application of the directive
- Published justification
Miro Analytical AG is the only vendor that can provide an analytical solution that enables the simultaneous measurement of all the major greenhouse gases (CO2, CH4, N2O and CO) and air quality pollutants (NO, NO2, SO2 and NH3) with the required high precision and broad amount fraction ranges that are required.
Unique features include:
- The ability to measure all these compounds simultaneously due to a unique dual colour laser system.
- The achieved precisions with this analyser match or exceed other optical spectrometers that are typically measuring one or two compounds.
- The range of amount fractions that are possible to measure exceed those typical of other optical spectrometers that are measuring one or two compounds.
Commercial outcome & contract awards
Published awards, contracts and bid statistics, not a forecast| Awarded to | Value | Date | Status |
|---|---|---|---|
| Miro Analytical AGSubcontracting: declared | CHF 205,274 † | — | Awarded |
Contracts
Quantum Cascade Laser Analyser for Greenhouse Gases and Air Pollutants · Miro Analytical AG
- Signed
- 28 Jul 2023
- Value
- CHF 205,274
- Status
- active
Contracting authority & participating public bodies
Dispute resolution & review bodies
- NPL Management LtdReview body
Bidders & participating suppliers
1 commercial party on the notice- Miro Analytical AGAwarded supplier
Official documents & specifications
Official links; attachments are not copiedNo document links were published on the notice.